Title :
Test of the HISPET prototype
Author :
Del Guerra, A. ; Bandettini, A. ; Conti, M. ; De Pascalis, G. ; Maiano, P. ; Perez Mendez, V. ; Rizzo, C.
Author_Institution :
Dipartimento di Sci. Fisiche, Napoli Univ., Italy
fDate :
4/1/1990 12:00:00 AM
Abstract :
The test of a small version of the high spatial resolution positron emission tomograph (HISPET) is presented. The HISPET prototype consists of two 50-cm×50-cm multiwire proportional chambers with a 2-cm-thick lead-glass drift space converter. A CAMAC acquisition system connected to a MAC II computer has been set up. Experimental data with a 22Na pointlike source have been reconstructed with a 3-D filtered backprojection algorithm. A spatial resolution better than 9 mm (full width at half maximum) and a quantitation capability at ±10% have been measured
Keywords :
CAMAC; computerised tomography; data acquisition; medical computing; radioisotope scanning and imaging; CAMAC acquisition system; HISPET; MAC II computer; PET; drift space converter; filtered backprojection algorithm; high spatial resolution positron emission tomograph; multiwire proportional chambers; positron emission tomography; quantitation capability; spatial resolution; Anodes; CAMAC; Delay lines; Glass; Laboratories; Positrons; Prototypes; Spatial resolution; Testing; USA Councils;
Journal_Title :
Nuclear Science, IEEE Transactions on