DocumentCode :
1049165
Title :
A tritium-in-air monitor with compensation and additional recording of α-, β-, and γ-backgrounds
Author :
Aoyama, Takahiko
Author_Institution :
Dept. of Nucl. Eng., Nagoya Univ., Japan
Volume :
37
Issue :
2
fYear :
1990
fDate :
4/1/1990 12:00:00 AM
Firstpage :
885
Lastpage :
891
Abstract :
The development of a computer-controlled tritium-in-air monitoring using flow-through air proportional counters as the detector is described. Combining automatic compensation for β- and γ-backgrounds with α-background discrimination, the monitor can detect tritium activity separately from other radioactive gases. The computer calculates and prints out not only tritium concentration in air but also dose equivalent rate by external γ-rays and/or activity concentrations of other radioactive gases than tritium in real time. Since the monitor has a wide range of 0.005-5000 Bq/cm3 (6 decades) and a short response time of several minutes, it can be used effectively to monitor tritium in air in fission and fusion reactor environments
Keywords :
proportional counters; radioactivity measurement; radioactivity measuring apparatus; tritium; T2; automatic compensation; dose equivalent rate; flow-through air proportional counters; radioactive gases; Automatic control; Computerized monitoring; Counting circuits; Fusion reactors; Gases; Inductors; Proportional control; Pulse circuits; Radiation detectors; Wire;