DocumentCode
105288
Title
Noncontact ECG Recording System With Real Time Capacitance Measurement for Motion Artifact Reduction
Author
Torfs, Tom ; Yun-Hsuan Chen ; Hyejung Kim ; Yazicioglu, Refet Firat
Author_Institution
Body Area Networks Dept., imec, Leuven, Belgium
Volume
8
Issue
5
fYear
2014
fDate
Oct. 2014
Firstpage
617
Lastpage
625
Abstract
A system for noncontact ECG recording is proposed that measures the real time electrode-body capacitance concurrently with the ECG as a reference signal for motion artifact reduction. Simultaneous recordings of these two signals from the human body in the presence of electrode motion artifacts are shown and an adaptive least-mean-squares (LMS) filtering algorithm run on these signals is demonstrated to be able to reduce the severity of certain types of electrode motion artifacts.
Keywords
bioelectric potentials; biomedical electrodes; capacitance; capacitance measurement; electrocardiography; least mean squares methods; medical signal processing; adaptive least-mean-squares filtering algorithm; motion artifact reduction; noncontact ECG recording system; real time electrode-body capacitance measurement; Capacitance; Capacitance measurement; Couplings; Electrocardiography; Electrodes; Impedance; Real-time systems; Adaptive filters; biomedical electrodes; capacitive sensors; electrocardiography; wearable sensors;
fLanguage
English
Journal_Title
Biomedical Circuits and Systems, IEEE Transactions on
Publisher
ieee
ISSN
1932-4545
Type
jour
DOI
10.1109/TBCAS.2014.2359053
Filename
6920090
Link To Document