• DocumentCode
    105729
  • Title

    High Field Insert Demonstrator Design, Manufacturing, and Tests of the Iseult Whole Body 11.75 T MRI Magnet

  • Author

    Schild, Thierry ; Bermond, S. ; Bredy, Philippe ; Donati, Andre ; Dubois, Olivier ; Gheller, J.M. ; Goc, J.J. ; Guillard, J.C. ; Lannou, Herve ; Leboeuf, R. ; Medioni, D. ; Nunio, Francois ; Molinie, Frederic ; Scola, Loris ; Sinanna, Armand ; Stepanov, V

  • Author_Institution
    Irfu, CEA/Saclay, Gif-sur-Yvette, France
  • Volume
    24
  • Issue
    3
  • fYear
    2014
  • fDate
    Jun-14
  • Firstpage
    1
  • Lastpage
    5
  • Abstract
    A neuroscience research center with very high field magnet resonance imaging (MRI) equipment has been opened in November 2006 in the Neurospin site of French Atomic Energy and Alternative Energies Commission (CEA, Saclay, France). One of the imaging systems, the so-called Iseult project, will require a whole body 11.75 T MRI magnet with a 900-mm warm bore. The coil is made of a niobium-titanium conductor cooled by a He II bath at 1.8 K, permanently connected to a cryoplant. The main coil is made of a stack of 170 double pancakes submitted to a peak field up to 12 T. A demonstrator made of six reduced double pancakes using the conductor developed for this project has been designed, manufactured, and tested at CEA/Saclay. The objective was to demonstrate that the Iseult main coil winding pack is able to sustain the high stress level calculated, 170 MPa azimuthally and 110 MPa radially. This demonstrator has been successfully energized up to 6000 A in a background field. A maximum azimuthal stress of 225 MPa has been reached, much higher than the nominal Iseult value. This paper presents the design, the manufacturing, and the cryogenics test results of this demonstrator.
  • Keywords
    biomedical MRI; biomedical equipment; conductors (electric); cryogenics; superconducting coils; Iseult main coil winding pack; Iseult whole body MRI magnet; azimuthal stress; bore; cryogenics test; cryoplant; double pancakes; high field insert demonstrator design; magnet resonance imaging; magnetic flux density 11.75 T; neuroscience research center; niobium-titanium conductor; pressure 110 MPa; pressure 225 MPa; radial stress; size 900 mm; temperature 1.8 K; Conductors; Magnetic resonance imaging; Magnetic tunneling; Magnetomechanical effects; Stress; Stress measurement; Superconducting magnets; Magnetic resonance imaging (MRI); niobium titanium; superconducting coil;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/TASC.2013.2291111
  • Filename
    6672006