• DocumentCode
    1057706
  • Title

    The non-proportional response of single crystalline oxide scintillators

  • Author

    Sysoeva, E.P. ; Zelenskaya, O.V. ; Sysoeva, E.V.

  • Author_Institution
    Inst. for Single Crystals, Kharkov, Ukraine
  • Volume
    43
  • Issue
    3
  • fYear
    1996
  • fDate
    6/1/1996 12:00:00 AM
  • Firstpage
    1282
  • Lastpage
    1283
  • Abstract
    The energy dependence of the response of single crystalline oxide scintillators (BGO, CWO and GSO(Ce)) was studied over the wide energy range from 6 to 1332 keV. The non-proportionality was defined as the difference between the specific response value at the given energy and the one at 661.6 keV. It is found that a considerable (up to 35%) non-proportionality of the response exists for the all investigated scintillators in the low energy region from 6 to 100 keV, while as in the energy range from 100 to 1332 keV the scintillator´s response is proportional to the gamma-rays energy with the accuracy equal to ±1%
  • Keywords
    gamma-ray detection; solid scintillation detectors; 6 to 100 keV; 6 to 1332 keV; 661.6 keV; BGO; CWO; GSO; energy dependence; gamma-ray energy; nonproportional response; single crystalline oxide scintillators; specific response value; Absorption; Bismuth; Cadmium compounds; Chemicals; Crystallization; Crystals; Energy resolution; Germanium; Tungsten; X-rays;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.506678
  • Filename