DocumentCode :
105778
Title :
Comparison of Analog and Digital Readout Electronics on the Processing of Charge-Sharing Events in Pixelated CdZnTe Detectors
Author :
Bolotnikov, A.E. ; Bale, D. ; Butcher, J. ; Camarda, G.S. ; Cui, Yan ; De Geronimo, G. ; Fried, J. ; Hossain, Abrar ; Kim, Ki Hyun ; Marshall, Matthew ; Soldner, S. ; Petryk, Matthew ; Prokesch, M. ; Vernon, E. ; Yang, Guo-Min ; James, Ralph B.
Author_Institution :
Brookhaven Nat. Lab., Upton, NY, USA
Volume :
61
Issue :
2
fYear :
2014
fDate :
Apr-14
Firstpage :
787
Lastpage :
792
Abstract :
We present our results from testing 15×15×10 mm3 CdZnTe pixelated detectors using a readout system based on the H3D ASIC. Data obtained with an uncollimated 137Cs source helped reveal details of the operational principle of such devices, and how the pulse-processing electronics may influence their performance. The responses of individual pixels were compared by using two different pulse processing approaches based on the analog shaping (H3D ASIC) and fitting of the waveforms as read out from hybrid preamplifiers.
Keywords :
application specific integrated circuits; nuclear electronics; position sensitive particle detectors; radioactive sources; readout electronics; semiconductor counters; 3D position sensitive detector; H3D ASIC; analog readout electronics; charge-sharing event processing; device operational principle; digital readout electronics; hybrid preamplifiers; pixelated CdZnTe detectors; pulse-processing electronics; uncollimated 137Cs source; Anodes; Application specific integrated circuits; Correlation; Detectors; Image quality; Readout electronics; Transient analysis; ASICs; CZT detectors; CdZnTe; crystal defects;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2014.2299493