Abstract :
An optically sensed isothermal differential thermal converter has been reduced to practice using a commercially available optical infrared pyrometer. Eliminating the conventional bimetallic thermocouples avoids draining thermal energy from the heater, reduces the Thomson effect error, and minimizes shunt paths along the heater which bypass higher frequency signal components. The optically sensed thermal element provides the potential for greater accuracy over a wide frequency range, faster response time, and improved dynamic range over conventional wire or planar thermal elements using Seebeck, thermoresistive or semiconductor temperature sensing
Keywords :
infrared detectors; pyrometers; temperature measurement; AC-DC conversion; Thomson effect error; dynamic range; optical infrared pyrometer; optically sensed isothermal differential thermal converter; optically sensed thermal element; response time; shunt paths; Dynamic range; Frequency; Isothermal processes; Optical sensors; Resistance heating; Shunt (electrical); Temperature sensors; Thermal conductivity; Thermoresistivity; Wire;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on