Title :
A 10 V calibrating system using an automated voltage divider and a 1 V Josephson array
Author :
Sakamoto, Yasuhiko ; Endo, Tadashi ; Sakuraba, Toshiaki
Author_Institution :
Electrotech. Lab., Ibaraki, Japan
fDate :
4/1/1993 12:00:00 AM
Abstract :
A 10-V calibrating system is developed by combining an automated voltage divider and a 1-V Josephson junction array voltage standard. The system can calibrate the 10 V output of a Zener voltage reference with an uncertainty of 0.02 ppm
Keywords :
Josephson effect; calibration; measurement standards; voltage dividers; voltage measurement; 10 V; Josephson array; Zener voltage reference; automated voltage divider; Calibration; Circuits; Electrical resistance measurement; Helium; Josephson junctions; Linearity; Resistors; Standards development; Uncertainty; Voltage;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on