  • Abstract
    This paper presents a method that detects and corrects alterations of 1, exposure; 2, out-of-focus; and 3, Gaussian noise affecting, contemporaneously, the images that are acquired in flow cytometer measurement devices. These alterations reduce image quality and interfere with correct micronucleus (MN) detection in a lymphocyte. The objectives of the proposed correction are given as follows: 1, to correctly process the image with the pattern-matching algorithm in order to detect the MN in human lymphocytes; 2, to minimize doubtful detections; and 3, to enhance the confidence that, in rejected images, MNs cannot be detected. Numerical and experimental tests confirm the validity of the proposed correction method and permit the evaluation of the upper and lower bounds of the admissible variation range of each alteration.
  • Keywords
    Gaussian noise; blood; blood flow measurement; cellular biophysics; image matching; medical image processing; Gaussian noise; contemporaneous alteration; flow cytometer measurement device; human lymphocyte; micronucleus image detection; pattern-matching algorithm; preprocessing correction; Fluid flow measurement; Focusing; Gaussian noise; Humans; Image enhancement; Image processing; Image quality; Noise measurement; Pixel; Testing; Exposure; Gaussian noise; flow cytometer; image enhancement; image processing; out of focus;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2007.900160
  • Filename
    4277009