Title :
Eight Tesla Superconducting Solenoid Magnet for the Magnetic Circular Dichroism Study
Author :
Chang, C.H. ; Hwang, C.S. ; Lin, F.Y. ; Huang, M.H. ; Chen, H.H. ; Jan, J.C. ; Yang, C.S. ; Lin, H.J.
Author_Institution :
Inst. of Nat. Synchrotron Radiat. Res. Center, Hsinchu
fDate :
6/1/2007 12:00:00 AM
Abstract :
This study describes the design and testing of a superconducting solenoid magnet for soft X-ray magnetic circular dichroism (SXMCD) experiments at the National Synchrotron Radiation Research Center (NSRRC). The superconducting solenoid magnet provides an uniform field of up to 8 T in a bore of 55 mm in diameter and 0.5 m long. The superconducting solenoid is operated in a 4.5-K liquid helium bath for reducing the vibration level during the experiments. The magnet and its cryostat are optimized to prevent magnet quenching and lower liquid helium consumption. The magnet quench test and magnetic field measurements are also performed.
Keywords :
cryostats; magnetic circular dichroism; superconducting magnets; cryostat; liquid helium bath; magnet quench test; magnetic field measurements; magnetic flux density 8 T; soft X-ray magnetic circular dichroism experiments; superconducting solenoid magnet; temperature 4.5 K; Boring; High temperature superconductors; Magnetic field measurement; Soft magnetic materials; Solenoids; Superconducting coils; Superconducting filaments and wires; Superconducting magnets; Synchrotron radiation; Testing; Magnetic circular dichroism; solenoid; superconducting;
Journal_Title :
Applied Superconductivity, IEEE Transactions on
DOI :
10.1109/TASC.2007.897806