Title
Variable gain X-ray image intensifier tube
Author
Carlson, Allan I. ; Singer, Barry M.
Author_Institution
Philips Laboratories, Briarcliff Manor, NY
Volume
26
Issue
11
fYear
1979
fDate
11/1/1979 12:00:00 AM
Firstpage
1711
Lastpage
1717
Abstract
An X-ray image intensifier tube with a variable gain SIT (silicon intensified target) target has been developed. The X-ray SIT tube has a gain control range of 300 to 1, low noise, 1500 TVL/ diameter limiting resolution, and picture quality comparable to that obtained with optically coupled X-ray intensifier-vidicon combinations. The variable internal gain of the X-ray SIT tube makes it possible to produce an electronic radiographic imaging system that can be used for both low-exposure-rate fluoroscopic examinations and high-exposure-rate, short-duration fluorographic recordings. Design considerations for the X-ray SIT target are presented, and a detailed analysis of the target noise is given. The measured characteristics of 6- and 9-in input diameter X-ray SIT tubes are summarized.
Keywords
Electron tubes; Gain control; High-resolution imaging; Image intensifiers; Optical coupling; Optical noise; Optical recording; Radiography; Silicon; X-ray imaging;