• DocumentCode
    1066899
  • Title

    Integrated signal conditioning for silicon pressure sensors

  • Author

    Borky, John M. ; Wise, Kensall D.

  • Author_Institution
    Air Force Institute of Technology, Wright-Patterson Air Force Base, Dayton, OH
  • Volume
    26
  • Issue
    12
  • fYear
    1979
  • fDate
    12/1/1979 12:00:00 AM
  • Firstpage
    1906
  • Lastpage
    1910
  • Abstract
    A silicon-diaphragm pressure sensor containing on-chip circuitry for signal conditioning has been designed for applications in biomedicine. The device requires no off-chip components and only two external leads, making it suitable for use with multisensor catheters having diameters less than 1.5 mm. The output is a temperature-compensated high-frequency FM analog representation of the applied pressure signal. The device exhibits 1-percent resolution and accuracy over the physiological pressure and temperature ranges. Circuit techniques applicable to a variety of active sensors are descried.
  • Keywords
    Biomedical measurements; Biosensors; Catheters; Circuits; Instruments; Sensor phenomena and characterization; Signal design; Silicon; Temperature distribution; Temperature sensors;
  • fLanguage
    English
  • Journal_Title
    Electron Devices, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9383
  • Type

    jour

  • DOI
    10.1109/T-ED.1979.19794
  • Filename
    1480372