Abstract :
In circumstances where it is not possible to make perfectly persistent joints, it is not necessary to rework all the joints as a simple flux pump injector could be built with the magnet to stabilize the decay. This is particularly important in the case of NMR magnets where the magnetic field decay is of order 0.01 ppm/hr. This paper describes a flux pump which is used in the first instance to energize a coil to a very high current. In a second stage, the flux pump was used to periodically inject a small incremental current to the magnet and thus stabilize its decay.
Keywords :
NMR spectroscopy; superconducting magnets; superconducting tapes; NMR magnets; current injection; flux injector; flux pump; magnetic field decay; nuclear magnetic resonance magnet; persistent joints; Boring; Circuits; Cryogenics; High temperature superconductors; Magnetic fields; Nuclear magnetic resonance; Superconducting coils; Superconducting filaments and wires; Superconducting magnets; Switches; Decay; flux pumping; magnetic fields; nuclear magnetic resonance spectroscopy;
Journal_Title :
Applied Superconductivity, IEEE Transactions on