Title :
Anthropomorphic cardiac ultrasound phantom
Author :
Smith, Stephen W. ; Rinaldi, Jean E.
Author_Institution :
Center for Devices & Radiol. Health, Food & Drug Adm., Rockville, MD, USA
Abstract :
A phantom which simulates the human cardiac anatomy for applications in ultrasound imaging, ultrasound Doppler, and color-flow Doppler imaging is described. The phantom consists of a polymer left ventricle which includes a prosthetic mitral and aortic valve and is connected to a mock circulatory loop. Aerated tap water serves as a blood simulating fluid and ultrasound contrast medium within the circulatory loop. The left ventricle is housed in a Lexan ultrasound visualization chamber which includes ultrasound viewing ports and acoustic absorbers. A piston pump connected to the visualization chamber by a single port pumps degassed water within the chamber which in turn pumps the left ventricle. Real-time ultrasound images and Doppler studies measure flow patterns through the valves and within the left ventricle.
Keywords :
biomedical equipment; biomedical ultrasonics; cardiology; Lexan ultrasound visualization chamber; aerated tap water; anthropomorphic cardiac ultrasound phantom; aortic valve; color-flow Doppler imaging; flow patterns; human cardiac anatomy simulation; mock circulatory loop; piston pump; polymer left ventricle; ultrasound Doppler; ultrasound contrast medium; ultrasound imaging; Anatomy; Anthropomorphism; Humans; Imaging phantoms; Polymers; Prosthetics; Pumps; Ultrasonic imaging; Valves; Visualization; Heart; Heart Ventricles; Humans; Models, Structural; Ultrasonography;
Journal_Title :
Biomedical Engineering, IEEE Transactions on