Title :
Fiberglass needle electrodes for transmural cardiac mapping
Author :
Rogers, Jack M. ; Melnick, Sharon B. ; Huang, Jian
Author_Institution :
Dept. of Biomed. Eng., Univ. of Alabama, Birmingham, AL, USA
Abstract :
We developed a new method for fabricating plunge needle electrodes for use in cardiac mapping. The needles have 12 electrodes with 1-mm spacing, are 0.5 mm in diameter, and are fabricated from fiberglass reinforced epoxy. They are stiff enough to be easily inserted into beating hearts and durable enough to be reused many times. We found that these new needles elicit smaller, more quickly resolving injury potentials, and when inserted in a row with 2-mm spacing, disrupt ventricular fibrillation activation patterns less than traditional steel needles.
Keywords :
biomedical electrodes; electrocardiography; glass fibre reinforced plastics; 0.5 mm; 1 mm; 2 mm; arrhythmia mechanisms; beating hearts; fiberglass needle electrodes; fiberglass reinforced epoxy; injury potentials; plunge needle electrodes; transmural cardiac mapping; ventricular fibrillation activation patterns; Electrodes; Fabrication; Fibrillation; Glass; Heart; Needles; Silver; Steel; Wires; Yarn; Animals; Body Surface Potential Mapping; Electrodes; Electrophysiology; Equipment Design; Equipment Failure Analysis; Glass; Heart Ventricles; Needles; Quality Control; Sensitivity and Specificity; Swine; Ventricular Fibrillation;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.2002.805483