Title :
Real-Time Intravascular Shear Stress in the Rabbit Abdominal Aorta
Author :
Ai, Lisong ; Yu, Hongyu ; Dai, Wangde ; Hale, Sharon L. ; Kloner, Robert A. ; Hsiai, Tzung K.
Author_Institution :
Dept. of Biomed. Eng. & Cardiovascular Med., Univ. of Southern California, Los Angeles, CA
fDate :
6/1/2009 12:00:00 AM
Abstract :
Fluid shear stress is intimately linked with the biological activities of vascular cells. A flexible microelectromechanical system (MEMS) sensor was developed to assess spatial- and temporal-varying components of intravascular shear stress (ISS) in the abdominal aorta of adult New Zealand white (NZW) rabbits. Real-time ISS (ISSreal-time) was analyzed in comparison with computational fluid dynamics (CFD) simulations for wall shear stress (WSS). Three-dimensional abdominal arterial geometry and mesh were created using the GAMBIT software. Simulation of arterial flow profiles was established by FLUENT. The Navier-Stokes equations were solved for non-Newtonian blood flow. The coaxial-wire-based MEMS sensor was deployed into the abdominal arteries of rabbits via a femoral artery cutdown. Based on the CFD analysis, the entrance length of the sensor on the coaxial wire (0.4 mm in diameter) was less than 10 mm. Three-dimensional fluoroscope and contrast dye allowed for visualization of the positions of the sensor and ratios of vessel to coaxial wire diameters. Doppler ultrasound provided the velocity profiles for the CFD boundary conditions. If the coaxial wire were positioned at the center of vessel, the CFD analysis revealed a mean ISS value of 31.1 with a systolic peak at 102.8 dyn ldr cm-2. The mean WSS was computed to be 10.1 dyn ldr cm-2 with a systolic peak at 33.2 dynldrcm-2, and the introduction of coaxial wire increased the mean WSS by 5.4 dyn ldr cm-2 and systolic peak by 18.0 dyn ldr cm-2. Experimentally, the mean ISS was 11.9 dyn ldr cm-2 with a systolic peak at 47.0 dyn ldr cm-2. The waveform of experimental ISS was similar to that of CFD solution with a 30.2% difference in mean and 8.9% in peak systolic shear stress. Despite the difference between CD and experimental results, the flexible coaxial-wire-based MEMS sensors provided a possibility to assess real-time ISS in the- - abdominal aorta of NZW rabbits.
Keywords :
Navier-Stokes equations; bioMEMS; biomedical ultrasonics; blood flow measurement; blood vessels; bone; cellular biophysics; computational fluid dynamics; diagnostic radiography; dyes; flow visualisation; medical computing; microsensors; muscle; Doppler ultrasound blood flow measurement; GAMBIT software; coaxial-wire-based MEMS sensor; computational fluid dynamics; dye visualization; microelectromechanical system; nonNewtonian blood flow; rabbit abdominal aorta; real-time intravascular cell fluid shear stress; temporal-spatial varying component; three-dimensional fluoroscope; Abdomen; Arteries; Cells (biology); Coaxial components; Computational fluid dynamics; Computational modeling; Micromechanical devices; Rabbits; Stress; Wire; Computational fluid dynamics (CFD); intravascular shear stress (ISS); microelectromechanical system (MEMS) sensors; Animals; Aorta, Abdominal; Blood Flow Velocity; Computer Simulation; Equipment Design; Hemorheology; Male; Micro-Electrical-Mechanical Systems; Prostheses and Implants; Pulsatile Flow; Rabbits; Shear Strength; Software; Stress, Mechanical; Ultrasonography, Doppler;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.2009.2013455