• DocumentCode
    1071206
  • Title

    A Digital Flux Injector Operated With a 317-MHz NMR Magnet

  • Author

    Yao, Weijun ; Kim, Woo-Seok ; Hahn, Seungyong ; BascuHán, J. ; Lee, Hai-Gun ; Iwasa, Yukikazu

  • Author_Institution
    MIT, Cambridge
  • Volume
    17
  • Issue
    2
  • fYear
    2007
  • fDate
    6/1/2007 12:00:00 AM
  • Firstpage
    1450
  • Lastpage
    1453
  • Abstract
    The paper reports results of the operation of a digital flux injector coupled to a 317-MHz NMR magnet operating at 4.2 K in persistent mode. The experiments have demonstrated that it is possible to periodically inject metered quantities of flux into a persistent-mode superconducting magnet to either compensate for a small field decay or to shift the operating frequency upward or downward without having to switch on the main power supply.
  • Keywords
    nuclear magnetic resonance; superconducting magnets; superconducting microwave devices; NMR magnet; digital flux injector; frequency 317 MHz; persistent-mode superconducting magnet; power supply; Couplings; High temperature superconductors; Magnetic flux; Nuclear magnetic resonance; Power supplies; Pulse transformers; Superconducting coils; Superconducting magnets; Superconductivity; Switches; Flux transformer; HTS/LTS magnet; NMR magnet; persistent mode;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/TASC.2007.898025
  • Filename
    4277827