• DocumentCode
    1071547
  • Title

    Three-dimensional optimization of correction iron pieces for open high field MRI system

  • Author

    Kakugawa, S. ; Hino, N. ; Komura, A. ; Kitamura, M. ; Takeshima, H. ; Yatsuo, T. ; Tazaki, H.

  • Author_Institution
    Hitachi Res. Lab., Hitachi Ltd., Ibaraki, Japan
  • Volume
    14
  • Issue
    2
  • fYear
    2004
  • fDate
    6/1/2004 12:00:00 AM
  • Firstpage
    1624
  • Lastpage
    1627
  • Abstract
    We have developed a new magnet for open high field (0.7 T) MRI systems. In order to homogenize magnetic fields in the field of view, an optimization method to determine initial configurations of large correction iron pieces was developed. In our method, we combined linear programming (LP) and 3D magnetic FEA and, using analytic formulae for LP, we computed field harmonics due to the volume of magnetized iron.
  • Keywords
    finite element analysis; linear programming; magnetic fields; magnetic resonance imaging; magnetisation; superconducting magnets; 0.7 T; 3D magnetic FEA; MRI magnet; analytic formulae; correction iron pieces; field harmonics; linear programming; magnetic fields homogeneity; magnetized iron; open high field MRI system; Iron; Linear programming; Magnetic analysis; Magnetic fields; Magnetic moments; Magnetic resonance imaging; Magnetic shielding; Magnetization; Optimization methods; Shape; Correction iron pieces; linear programming; optimization; uniform magnetic field;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/TASC.2004.831019
  • Filename
    1325114