DocumentCode :
1072034
Title :
Development of temperature fixed points applicable to the calibration of thermometers in magnetic fields
Author :
Nara, Koichi
Author_Institution :
NMIJ/AIST, Ibaraki, Japan
Volume :
14
Issue :
2
fYear :
2004
fDate :
6/1/2004 12:00:00 AM
Firstpage :
1798
Lastpage :
8801
Abstract :
Two temperature fixed points, (H2O and Ar) were developed to be used under magnetic fields. They serve as ideal tools to investigate the magnetic field dependence of temperature sensors. The sensors studied are Cernox™, a platinum resistance thermometer, and a thermistor. The magnetic field effect on Cernox was found to be not larger than 5 mK under 15 T at the triple point of water. The applicability of Kohler´s rule was evaluated by a PRT and was found to be valid within 3%. The magnetic field effect on the thermistor studied was found to be 15 mK at 15 T.
Keywords :
calibration; magnetic field effects; magnetic fields; resistance thermometers; temperature measurement; thermistors; thermometers; 15 T; 15 mK; Cernox™; Kohler rule; magnetic fields; platinum resistance thermometer; temperature fixed points; temperature sensors; thermistor; thermometer calibration; Calibration; Capacitance; Glass; Magnetic field measurement; Magnetic fields; Magnetic sensors; Temperature control; Temperature sensors; Thermal sensors; Thermistors; Calibration; Cernox; magnetic field effect; platinum resistance thermometer; temperature fixed point; temperature measurement; thermistor;
fLanguage :
English
Journal_Title :
Applied Superconductivity, IEEE Transactions on
Publisher :
ieee
ISSN :
1051-8223
Type :
jour
DOI :
10.1109/TASC.2004.831115