DocumentCode
1072327
Title
Point source resolution performance of a germanium Compton camera
Author
Haskins, P.S. ; McKisson, J.E. ; Neelands, K.C. ; Henderson, D.P., Jr.
Author_Institution
Radiatian Technol. Inc., Alachua, FL, USA
Volume
43
Issue
3
fYear
1996
fDate
6/1/1996 12:00:00 AM
Firstpage
1832
Lastpage
1836
Abstract
One of the critical performance parameters for a Compton camera imager is the point source location accuracy. This paper describes analysis of both experimental and simulated data to evaluate the point source resolving power of a germanium Compton camera imager. The experimental data are taken from a set of experiments performed in 1993 at Naval Research Laboratory. Subsequent design studies have utilized Monte Carlo simulations to predict the performance parameters for a number of detector geometries. Presented in this paper are the results from one of those configurations
Keywords
Compton effect; Monte Carlo methods; cameras; germanium radiation detectors; Compton camera imager; Ge; Monte Carlo simulations; germanium Compton camera; point source location accuracy; point source resolution performance; Cameras; Detectors; Germanium; Image reconstruction; Laboratories; Position measurement; Radiofrequency interference; Scattering parameters; Surface reconstruction; Uncertainty;
fLanguage
English