• DocumentCode
    1073211
  • Title

    Digital mammography at the Trieste synchrotron light source

  • Author

    Arfelli, F. ; Barbiellini, G. ; Bonvicin, V. ; Bravin, A. ; Cantatore, G. ; Castelli, E. ; Palma, L. Dalla ; Di Michiel, M. ; Longo, R. ; Olivo, A. ; Pani, S. ; Pontoni, D. ; Poropat, P. ; Prest, M. ; Rosei, R. ; Sessa, M. ; Tromba, G. ; Vacchi, A.

  • Author_Institution
    Dipartimento di Fisica, Trieste Univ., Italy
  • Volume
    43
  • Issue
    3
  • fYear
    1996
  • fDate
    6/1/1996 12:00:00 AM
  • Firstpage
    2061
  • Lastpage
    2067
  • Abstract
    The SYRMEP collaboration is developing a digital mammography project using a synchrotron radiation monochromatic X-ray beam and a silicon pixel detector. The combination of a monochromatic laminar beam with a high efficiency laminar detector, capable of single-photon counting, allows one to minimize the radiation dose delivered to the sample, while maximizing contrast resolution and dynamic range. The SYRMEP detector is a silicon microstrip device used in an innovative configuration in which radiation impinges on the side rather than on the surface of the chip and is therefore totally absorbed within the detector active volume. The high contrast resolution and spatial resolution (<100 microns) of the system have been demonstrated by producing images of details from an Ackermann Mammographic Phantom RMI 160 illuminated by a conventional X-ray tube
  • Keywords
    biomedical equipment; diagnostic radiography; silicon radiation detectors; synchrotron radiation; Ackermann Mammographic Phantom RMI 160; Si; Trieste synchrotron light source; breast cancer detection; conventional X-ray tube; detector active volume; digital mammography; high contrast resolution; high efficiency laminar detector; medical diagnostic imaging; medical instrumentation; silicon microstrip device; silicon pixel detector; single-photon counting; spatial resolution; synchrotron radiation monochromatic X-ray beam; totally absorbed radiation; Collaboration; Dynamic range; Light sources; Mammography; Radiation detectors; Silicon radiation detectors; Spatial resolution; Synchrotron radiation; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.507269
  • Filename
    507269