DocumentCode :
1073306
Title :
A 122-channel whole-cortex SQUID system for measuring the brain´s magnetic fields
Author :
Knuutila, Jukka E T ; Ahonen, Antti I. ; Hamalainen, Matti S. ; Kajola, Matti J. ; Laine, P. Petteri ; Lounasmaa, Olli V. ; Parkkonen, Lauri T. ; Simola, Juha T A ; Tesche, Claudia D.
Author_Institution :
Low Temp. Lab., Helsinski Univ. of Technol., Espoo, Finland
Volume :
29
Issue :
6
fYear :
1993
fDate :
11/1/1993 12:00:00 AM
Firstpage :
3315
Lastpage :
3320
Abstract :
A 122-channel neuromagnetometer with a helmet-shaped detector array covering the entire head allows simultaneous recording of magnetic fields over the whole cortex. The instrument has 122 planar first-order gradiometers in dual units at 61 measurement sites. The SQUIDs are directly coupled to the read-out electronics, with amplifier noise cancellation to eliminate the need for separate preamplifiers inside the magnetically shielded room. The authors analyze the performance of the device and compare it with traditional axial gradiometer arrays by considering signal-to-noise ratios, spatial sampling theory, confidence intervals for equivalent current dipole fits, and information-theoretical channel capacity. The analysis includes the fact that instrument noise is smaller than the background activity of the brain; the signal-to-noise ratio and the resolution of the planar array are in that case equal to or better than that of an axial array. The number of channels and their spacing are very suitable for neuromagnetic measurements
Keywords :
SQUIDs; biological techniques and instruments; biomagnetism; brain; magnetic field measurement; 122-channel whole-cortex SQUID system; amplifier noise cancellation; axial gradiometer arrays; brain research instrumentation; equivalent current dipole fits; helmet-shaped detector array; information-theoretical channel capacity; magnetically shielded room; neuromagnetic measurements; planar first-order gradiometers; preamplifiers; read-out electronics; signal-to-noise ratio; Detectors; Instruments; Magnetic analysis; Magnetic field measurement; Magnetic heads; Magnetic shielding; SQUIDs; Sensor arrays; Signal analysis; Signal to noise ratio;
fLanguage :
English
Journal_Title :
Magnetics, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9464
Type :
jour
DOI :
10.1109/20.281163
Filename :
281163
Link To Document :