• DocumentCode
    1075626
  • Title

    Highly Sensitive Measurement of Moisture Content Using HTS-SQUID

  • Author

    Tsukada, Keiji ; Kiwa, Toshihiko ; Tahara, Hideaki ; Miyake, Etsuro ; Yamada, Hironobu ; Tsukamoto, Akira ; Kandori, Akihiko

  • Author_Institution
    Grad. Sch. of Natural Sci. & Technol., Okayama Univ., Okayama, Japan
  • Volume
    19
  • Issue
    3
  • fYear
    2009
  • fDate
    6/1/2009 12:00:00 AM
  • Firstpage
    878
  • Lastpage
    881
  • Abstract
    We developed a highly sensitive magnetic system for measuring moisture content using an HTS-SQUID sensor. The system consists mainly of an exposure coil, a Dewar flask containing HTS-SQUID immersed in liquid nitrogen, a cancellation coil, a lock-in amplifier, magnetically shielded box, and FLL circuit. The cancellation coil was used to eliminate the applied magnetic field at the sensor position, and then a cancellation ratio of more than -40 dB was obtained. By the high performance, the subject can be exposed to a powerful magnetic field strength in the order of muT, and a weak magnetic signal of less than tens of pT from the subject can be detected. Using a magnetic vector analysis technique, more precise measurement of moisture content is realized.
  • Keywords
    SQUIDs; high-temperature superconductors; magnetic sensors; moisture measurement; Dewar flask; FLL circuit; HTS-SQUID sensor; cancellation coil; exposure coil; liquid nitrogen; lock-in amplifier; magnetic field strength; magnetic vector analysis technique; moisture measurement; sensitive magnetic sensor; sensitive magnetic system; Diamagnetism; HTS-SQUID; magnetic cancellation; moisture content;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/TASC.2009.2019659
  • Filename
    5075618