• DocumentCode
    1079545
  • Title

    Three-dimensional (3-D) reconstruction of the spine from a single X-ray image and prior vertebra models

  • Author

    Novosad, Justin ; Cheriet, Farida ; Petit, Yvan ; Labelle, Hubert

  • Author_Institution
    Ste.-Justine Hosp. Res. Center, Montreal, Que., Canada
  • Volume
    51
  • Issue
    9
  • fYear
    2004
  • Firstpage
    1628
  • Lastpage
    1639
  • Abstract
    The lateral bending test is routinely used by clinicians for the preoperative assessment of spinal mobility. The evaluation of bending motion is usually based on the qualitative analysis of a two-dimensional (2-D) antero-posterior X-ray image. The aim of this paper is to introduce a novel three-dimensional (3-D) reconstruction technique that is a prerequisite for the quantitative 3-D analysis of lateral bending motion. An algorithm was developed for the 3-D reconstruction of the spine from a single X-ray image. The X-ray is calibrated using a small calibration object and an explicit calibration algorithm. The information contained in the single X-ray is completed by registering a priori 3-D geometric models of individual vertebrae. Part of the error yielded by the 3-D/2-D registration is corrected by a vertebral alignment constraint that aims to minimize intervertebral dislocations. Three-dimensional models of 15 different scoliosis patients, obtained from a standard stereo-radiographic 3-D reconstruction, were used in simulation and validation experiments. Experimental results show that the new method is robust and accurate. With pessimistic levels of simulated noise, the average root mean square reconstruction error is 2.89 mm, which is appropriate for common clinical applications.
  • Keywords
    bending; biomechanics; bone; diagnostic radiography; diseases; image motion analysis; image reconstruction; image registration; medical image processing; a priori 3-D geometric models; bending motion; calibration; image registration; intervertebral dislocations; lateral bending test; scoliosis patients; single X-ray image; spinal mobility; stereo-radiographic 3-D reconstruction; three-dimensional spine reconstruction; two-dimensional antero-posterior X-ray image; vertebra models; Calibration; Image analysis; Image motion analysis; Image reconstruction; Motion analysis; Solid modeling; Testing; Three dimensional displays; Two dimensional displays; X-ray imaging; Algorithms; Artificial Intelligence; Calibration; Cluster Analysis; Computer Simulation; Imaging, Three-Dimensional; Information Storage and Retrieval; Models, Biological; Models, Statistical; Pattern Recognition, Automated; Phantoms, Imaging; Radiographic Image Enhancement; Radiographic Image Interpretation, Computer-Assisted; Reproducibility of Results; Scoliosis; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Spine; Subtraction Technique;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2004.827537
  • Filename
    1325823