DocumentCode :
1082928
Title :
Ultrasound Current Source Density Imaging
Author :
Olafsson, Ragnar ; Witte, Russell S. ; Huang, Sheng-Wen ; Donnell, Matthew O.
Author_Institution :
Dept. of Biomed. Eng., Univ. of Michigan, Ann Arbor, MI
Volume :
55
Issue :
7
fYear :
2008
fDate :
7/1/2008 12:00:00 AM
Firstpage :
1840
Lastpage :
1848
Abstract :
Surgery to correct severe heart arrhythmias usually requires detailed maps of the cardiac activation wave prior to ablation. The pinpoint electrical mapping procedure is laborious and limited by its spatial resolution (5-10 mm). We propose ultrasound current source density imaging (UCSDI), a direct 3-D imaging technique that potentially facilitates existing mapping procedures with superior spatial resolution. The technique is based on a pressure-induced change in resistivity known as the acoustoelectric (AE) effect, which is spatially confined to the ultrasound focus. AE-modulated voltage recordings are used to map and reconstruct current densities. In this preliminary study, we tested UCSDI under controlled conditions and compared it with conventional electrical mapping techniques. A 2-D dipole field was produced by a pair of electrodes in a bath of 0.9% NaCl solution. Boundary electrodes detected the AE signal while a 7.5-MHz focused ultrasound transducer was scanned across the bath. UCSDI located the current source and sink to within 1 mm of their actual positions. A future UCSDI system potentially provides real-time 3-D images of the cardiac activation wave coregistered with anatomical ultrasound and would greatly facilitate corrective procedures for heart abnormalities.
Keywords :
acoustoelectric effects; bioacoustics; bioelectric phenomena; biomedical ultrasonics; cardiology; diseases; medical image processing; surgery; 3D imaging technique; acoustoelectric effect; boundary electrodes; cardiac activation wave; focused ultrasound transducer; heart abnormality; mapping procedures; pinpoint electrical mapping procedure; resistivity; severe heart arrhythmias; spatial resolution; surgery; ultrasound current source density imaging; Conductivity; Current density; Electrodes; Focusing; Heart; Image reconstruction; Spatial resolution; Surgery; Ultrasonic imaging; Voltage; Acoustics; acoustics; acoustoelectric effect; bioelectric phenomena; electrocardiography; Algorithms; Animals; Body Surface Potential Mapping; Heart Conduction System; Humans; Image Interpretation, Computer-Assisted; Imaging, Three-Dimensional; Ultrasonography;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2008.919115
Filename :
4457869
Link To Document :
بازگشت