• DocumentCode
    1084268
  • Title

    Propagation of errors from the sensitivity image in list mode reconstruction

  • Author

    Qi, Jinyi ; Huesman, Ronald H.

  • Author_Institution
    Dept. of Nucl. Med. & Functional Imaging, Lawrence Berkeley Nat. Lab., CA, USA
  • Volume
    23
  • Issue
    9
  • fYear
    2004
  • Firstpage
    1094
  • Lastpage
    1099
  • Abstract
    List mode image reconstruction is attracting renewed attention. It eliminates the storage of empty sinogram bins. However, a single back projection of all LORs is still necessary for the pre-calculation of a sensitivity image. Since the detection sensitivity is dependent on the object attenuation and detector efficiency, it must be computed for each study. Exact computation of the sensitivity image can be a daunting task for modern scanners with huge numbers of LORs. Thus, some fast approximate calculation may be desirable. In this paper, we analyze the error propagation from the sensitivity image into the reconstructed image. The theoretical analysis is based on the fixed point condition of the list mode reconstruction. The nonnegativity constraint is modeled using the Kuhn-Tucker condition. With certain assumptions and the first-order Taylor series approximation, we derive a closed form expression for the error in the reconstructed image as a function of the error in the sensitivity image. The result shows that the error response is frequency-dependent and provides a simple expression for determining the required accuracy of the sensitivity image calculation. Computer simulations show that the theoretical results are in good agreement with the measured results.
  • Keywords
    error analysis; image reconstruction; medical image processing; positron emission tomography; Kuhn-Tucker condition; PET scanner; detector efficiency; error propagation; first-order Taylor series approximation; fixed point condition; frequency-dependent error response; list mode image reconstruction; nonnegativity constraint; object attenuation; sensitivity image; Attenuation; Computer errors; Computer simulation; Detectors; Error analysis; Frequency; Image analysis; Image reconstruction; Object detection; Taylor series; Algorithms; Computer Simulation; Data Interpretation, Statistical; Image Enhancement; Image Interpretation, Computer-Assisted; Information Storage and Retrieval; Likelihood Functions; Models, Theoretical; Reproducibility of Results; Sensitivity and Specificity;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2004.829333
  • Filename
    1327688