• DocumentCode
    1092510
  • Title

    Evaluation of Image Noise in Respiratory Gated PET

  • Author

    Detorie, Nicole C. ; Kesner, Adam L. ; Solberg, Timothy D. ; Dahlbom, Magnus

  • Author_Institution
    Dept. of Molecular & Med. Pharmacology, California Univ., Los Angeles, CA
  • Volume
    54
  • Issue
    1
  • fYear
    2007
  • Firstpage
    66
  • Lastpage
    70
  • Abstract
    The aim of this study was to quantify image noise and signal recovery in respiratory gated PET. A Jaszczak phantom filled with 18F was placed on a custom built motion platform. Different source to background activity ratios were used. An Anzai belt, a surface tension monitoring device, was strapped around the phantom to track the motion and to trigger the gated PET cycle. Data were acquired into 12 bins throughout one gating cycle. The binned data were also summed to produce image sets representing acquisitions with different numbers of gates, including a non-gated image set. The image noise was estimated using the bootstrap method. Images were generated from 100 sinogram replicates and reconstructed using ordered subsets-expectation maximization (OSEM), 4 iterations and 8 subsets. From the reconstructed image replicates, mean and standard deviation images were created, from which the average image signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of each sphere were calculated. The SNR and CNR were evaluated as a function of the number of gates. The SNR and CNR result in the expected Poisson limited correlation with the number of gates for the larger lesions. Because of the motion, the CNR calculated from the images produced with no or few gates is nearly a factor of 2 less than the expected value for the 3 smallest spheres. As the number of gates increases, the CNR correlates with the expected values. Beyond 6 gates, image noise dominates over any signal improvement, which is reflected in the low CNR values of the smallest spheres. The results of this study show that gating can provide improvement in signal recovery with minimal loss of CNR for small, moving lesions
  • Keywords
    expectation-maximisation algorithm; image denoising; image motion analysis; image reconstruction; lung; medical image processing; phantoms; pneumodynamics; positron emission tomography; 18F; CNR; Jaszczak phantom; SNR; contrast-to-noise ratio; image noise; image reconstruction; ordered subsets-expectation maximization; respiratory gated PET; respiratory motion; signal recovery; signal-to-noise ratio; standard deviation; surface tension monitoring device; Belts; Image generation; Image reconstruction; Imaging phantoms; Lesions; Monitoring; Positron emission tomography; Signal to noise ratio; Surface tension; Tracking; Gating; PET; image noise; respiratory motion;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2006.890014
  • Filename
    4089158