DocumentCode :
1093306
Title :
On detecting the presence of fetal R-wave using the moving averaged magnitude difference algorithm
Author :
Park, Y.C. ; Lee, K.Y. ; Youn, D.H. ; Kim, N.H. ; Kim, W.K. ; Park, S.H.
Author_Institution :
Yonsei Univ., Seoul, South Korea
Volume :
39
Issue :
8
fYear :
1992
Firstpage :
868
Lastpage :
871
Abstract :
An efficient algorithm for detecting the presence of a fetal QRS complex is presented. The proposed fetal QRS detection method computes the averaged magnitude of the difference between the fetal ECG signal and the reference signal to detect the fetal QRS event. The detected signal fetal QRS complexes are exponentially averaged to generate the template signal which can track the slowly varying shape of the fetal ECG signal. As an effort to obtain improved detection performances, two approaches of normalizing the fetal ECG signal and the template are considered.
Keywords :
electrocardiography; signal processing; exponential averaging; fetal ECG signal; fetal QRS complex; fetal R-wave; moving averaged magnitude difference algorithm; slowly varying signal shape; template signal; Acoustic noise; Detection algorithms; Electrocardiography; Event detection; Matched filters; Motion detection; Noise cancellation; Shape; Signal processing; Signal processing algorithms; Algorithms; Electrocardiography; Fetal Monitoring; Humans; Sensitivity and Specificity; Signal Processing, Computer-Assisted;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.148396
Filename :
148396
Link To Document :
بازگشت