DocumentCode :
1093683
Title :
Improvement of the Voltage Dependence of High-Voltage AC–DC Transfer Differences at the NMIJ
Author :
Fujiki, Hiroyuki
Author_Institution :
Nat. Inst. of Adv. Ind. Sci. & Technol., Nat. Metrol. Inst. of Japan, Tsukuba
Volume :
57
Issue :
9
fYear :
2008
Firstpage :
1992
Lastpage :
1997
Abstract :
A new AC-DC comparator system of 20-1000 V has been developed at the National Metrology Institute of Japan (NMIJ) for the calibration of ac-dc thermal voltage converters (TVCs), which are used as national ac-dc transfer standards. The ac-dc transfer differences of high-voltage transfer standards were evaluated by a traditional step-up procedure. The voltage dependence of the ac-dc transfer difference was observed in the earlier step-up procedures over 300 V, as reported by the latest international intercomparison. The experimental results for high-voltage TVCs with several different range resistors at the NMIJ suggest that the voltage dependence may primarily be caused by input connectors of range resistors and the change in the resistance value of the resistor and TVC modules due to heating from resistors. This paper describes the voltage dependence improvement of ac-dc transfer differences over 300 V at the NMIJ.
Keywords :
AC-DC power convertors; calibration; resistors; AC-DC comparator system; AC-DC thermal voltage converters; National Metrology Institute of Japan; calibration; high-voltage AC-DC transfer differences; international intercomparison; range resistor; voltage 20 V to 1000 V; AC–DC transfer difference; AC??DC transfer difference; ac voltage standard; range resistor; thermal voltage converters (TVCs); voltage dependence;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2008.917256
Filename :
4464112
Link To Document :
بازگشت