DocumentCode :
1097359
Title :
Comments, with reply, on "Monte Carlo simulation of BaF/sub 2/ detectors used in time-of-flight positron emission tomography" by A.N. Bice et al
Author :
Muzic, Raymond F., Jr. ; Nelson, A. Dennis ; Miraldi, Floro
Author_Institution :
Univ. Hospitals of Cleveland, OH, USA
Volume :
38
Issue :
2
fYear :
1991
fDate :
4/1/1991 12:00:00 AM
Firstpage :
891
Abstract :
The commenters point out an error in Fig.2 of the above-named work of A.N. Bice et al. (see ibid., vol.37, no.2, p.696-701, April 1990). The error arose because Bice et al. inadvertently used a Gaussian distribution with sigma equal to 35% of the photopeak of 0.179 MeV. In a reply. Bice et al. acknowledge the error.<>
Keywords :
barium compounds; radioisotope scanning and imaging; scintillation counters; BaF/sub 2/; Gaussian distribution; Monte Carlo simulation; time-of-flight positron emission tomography; Crystals; Detectors; Energy measurement; Energy resolution; Hospitals; Laboratories; Optical coupling; Radiology;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.289406
Filename :
289406