DocumentCode
1097477
Title
Deterioration of average evoked potential waveform due to asynchronous averaging and its compensation
Author
Nakamura, Masatoshi ; Nishida, Shigeto ; Shibasaki, Hiroshi
Author_Institution
Dept. of Electr. Eng., Saga Univ., Japan
Volume
38
Issue
3
fYear
1991
fDate
3/1/1991 12:00:00 AM
Firstpage
309
Lastpage
312
Abstract
The deterioration of the average due to the asynchronous sampling of evoked potential data is discussed. The asynchronous averaging in the frequency domain is analyzed, and a relationship between the fluctuating time and the degree of the deterioration is derived. By using this relationship, a required sampling rate is derived to ensure the accuracy of the averaging. In the case of coarse sampling, a compensation procedure for the asynchronous averaging is deduced, by which the compensated signal approaches the original signal as the number of averagings increases.
Keywords
bioelectric potentials; signal processing; asynchronous averaging; average evoked potential waveform; coarse sampling; compensated signal; fluctuating time; sampling rate; waveform deterioration; Biomedical measurements; Distortion measurement; Electric variables measurement; Electromyography; Frequency estimation; Noise measurement; Pollution measurement; Sampling methods; Signal to noise ratio; Silicon compounds; Evoked Potentials; Signal Processing, Computer-Assisted;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/10.133216
Filename
133216
Link To Document