Title :
The Micro Reading Mesh Chamber
Author :
Fernández, Ferran ; Ramos-Lerate, Inma ; Martínez, Juan Carlos ; Beltrán, David
Author_Institution :
Lab. de Llum de Sincrotro
Abstract :
Micro-pattern gas detectors (MPGD) could become suitable devices to carry out time-resolved X-ray diffraction experiments in the sub-millisecond time scale at synchrotron radiation facilities. Here we introduce a new MPGD design we have called micro reading mesh chamber (MRMC). It is a position-sensitive gas proportional counter that is built with advanced printed circuit board (PCB) technology. Its layout is based on a mesh formed by two planes of pick-up strips that is separated from the resistive anode by support pillars that set the multiplication region. The aim of the MRMC design has been to prevent the arising of dielectric breakdown and to avoid charging up. We present detailed simulation/optimization work and first experimental results.
Keywords :
X-ray detection; X-ray diffraction; position sensitive particle detectors; proportional counters; readout electronics; synchrotron radiation; MPGD; MRMC; Micro Reading Mesh Chamber; dielectric breakdown; micro-pattern gas detectors; position-sensitive gas proportional counter; printed circuit board technology; resistive anode; sub-millisecond time scale; synchrotron radiation; time-resolved X-ray diffraction experiments; Aging; Anodes; Breakdown voltage; Cathodes; Dielectric breakdown; Spatial resolution; Strips; Synchrotron radiation; X-ray detectors; X-ray diffraction; Gas detectors; X-ray detectors; position sensitive particle detectors;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2007.901229