DocumentCode :
1102013
Title :
Comparison of DC Voltage Standards using a 10:1 divider
Author :
De Souza, Heriguatí ; Slomovitz, Daniel
Volume :
4
Issue :
3
fYear :
2006
fDate :
5/1/2006 12:00:00 AM
Firstpage :
156
Lastpage :
159
Abstract :
A self-calibrated resistive divider to relate 1 V and 10 V levels of DC voltage standards (standard cells, Zener sources, Josephson systems) is presented. The system is based on a divider formed by ten equal resistors, in which the voltage drop on each resistor is measured as well as the total voltage drop. The ratio between the total voltage drop and the average voltage drop on each resistor is, theoretically, 10:1. This measuring system only requires of the divider and a multimeter of 8½ digits as auxiliary equipment. The measurement uncertainty, using this divider, is in the order of 1´10-7.
Keywords :
Instruments; Laboratories; Resistors; Voltage; Josephson; Resistive Divider; Standard Cell; Zener;
fLanguage :
English
Journal_Title :
Latin America Transactions, IEEE (Revista IEEE America Latina)
Publisher :
ieee
ISSN :
1548-0992
Type :
jour
DOI :
10.1109/TLA.2006.4472107
Filename :
4472107
Link To Document :
بازگشت