• DocumentCode
    1106241
  • Title

    Tooth-Click Control of a Hands-Free Computer Interface

  • Author

    Simpso, Tyler ; Broughton, Colin ; Gauthier, Michel J A ; Prochazka, Arthur

  • Author_Institution
    Centre for Neurosci., Alberta, Univ., Edmonton, AB
  • Volume
    55
  • Issue
    8
  • fYear
    2008
  • Firstpage
    2050
  • Lastpage
    2056
  • Abstract
    People with severe upper limb paralysis use devices that monitor head movements to control computer cursors. The three most common methods for producing mouse button clicks are dwell-time, sip-and-puff control, and voice-recognition. Here, we tested a new method in which small tooth-clicks were detected by an accelerometer contacting the side of the head. The resulting signals were paired with head tracking technology to provide combined cursor and button control. This system was compared with sip-and-puff control and dwell-time selection. A group of 17 people with disabilities and ten people without disabilities tested each system by producing mouse clicks as inputs to two software programs. Tooth-click/head-mouse control was much faster than dwell-time control and not quite as fast as sip-and-puff control, but it was more reliable and less cumbersome than the latter.
  • Keywords
    handicapped aids; user interfaces; accelerometer; dwell-time control; hands-free computer interface; head-mouse control; mouse button clicks; sip-and-puff control; tooth-click control; upper limb paralysis; voice-recognition; Accelerometers; Cameras; Computer interfaces; Computerized monitoring; Diseases; Head; Internet; Keyboards; Mice; Neuroscience; Spinal cord injury; Teeth; Tracking; Assistive technology (AT); assistive technology; computer interfaces; spinal cord injury; tooth-click; Acceleration; Adult; Auscultation; Computer Peripherals; Equipment Design; Equipment Failure Analysis; Female; Hand; Hemiplegia; Humans; Male; Middle Aged; Spinal Cord Injuries; Tooth; Transducers; User-Computer Interface;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2008.921161
  • Filename
    4473368