DocumentCode :
1110100
Title :
A frameless stereotaxic operating microscope for neurosurgery
Author :
Friets, Eric M. ; Strohbehn, John W. ; Hatch, John F. ; Robert, David W.
Author_Institution :
Thayer Sch. of Eng., Dartmouth Coll., Hanover, NH, USA
Volume :
36
Issue :
6
fYear :
1989
fDate :
6/1/1989 12:00:00 AM
Firstpage :
608
Lastpage :
617
Abstract :
The purpose of the frameless stereotaxic operating microscope is to display computed tomography (CT) or other image data in the operating microscope in the correct scale, orientation, and position without the use of a stereotaxic frame. A nonimaging ultrasonic rangefinder allows the position of the operating microscope and the position of the patient to be determined. Discrete fiducial points on the patient´s external anatomy are located in both image space and operating room space, linking the image data and the operating room. Physician-selected image information, e.g. tumor contours or guidance to predetermined targets, is projected through the optics of the operating microscope using a miniature cathode ray tube and a beam splitter. Projected images superpose the surgical field, reconstructed from image data to match the focal plane of the operating microscope. The algorithms on which the system is based are described, and the sources and effects of errors are discussed. The system´s performance is simulated, providing an estimate of accuracy. Two phantoms are used to measure accuracy experimentally. Clinical results and observations are given.
Keywords :
biomedical equipment; brain; optical microscopes; surgery; beam splitter; cathode ray tube; fiducial points; frameless stereotaxic operating microscope; neurosurgery; nonimaging ultrasonic rangefinder; phantoms; Anatomy; Cathode ray tubes; Computed tomography; Displays; Joining processes; Neoplasms; Neurosurgery; Optical beam splitting; Optical microscopy; Surgery; Algorithms; Humans; Image Processing, Computer-Assisted; Microscopy; Models, Structural; Neurosurgery; Stereotaxic Techniques; Tomography, X-Ray Computed;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.29455
Filename :
29455
Link To Document :
بازگشت