DocumentCode :
1115283
Title :
Body area network for wireless patient monitoring
Author :
Montón, E. ; Hernandez, J.F. ; Blasco, J.M. ; Hervé, T. ; Micallef, J. ; Grech, I. ; Brincat, A. ; Traver, V.
Author_Institution :
ITACA Inst., Polytech. Univ. of Valencia, Valencia
Volume :
2
Issue :
2
fYear :
2008
fDate :
2/1/2008 12:00:00 AM
Firstpage :
215
Lastpage :
222
Abstract :
Patient data monitoring is a key issue for health and disease management. The use of wireless sensors within a body area network (BAN) makes this task seamless and easy. A BAN system is presented, which allows the connectivity of a wide range of heterogeneous body sensors to a portable hub device that is connectable to external networks (IEEE 802.11, GPRS). This BAN is based on the use of Zigbee/IEEE 802.15.4 standard technology and off-the-shelf modules. It is currently being used at the European level for the detection and the prediction of the human physiological state in relation to wakefulness, fatigue, and stress applications in which users carrying out daily activities are monitored in an unobtrusive and comfortable way. Characterised by its low power consumption, low cost, and ability to connect a wide range of heterogeneous sensors, this system can substantially improve the performance of different services, especially those that are health related.
Keywords :
IEEE standards; biomedical communication; health care; low-power electronics; packet radio networks; patient monitoring; wireless LAN; wireless sensor networks; GPRS; IEEE 802.11; Zigbee/IEEE 802.15.4 standard technology; body area network; disease management; health management; heterogeneous body sensors; low power consumption; off-the-shelf modules; patient data monitoring; portable hub device; wireless patient monitoring; wireless sensors;
fLanguage :
English
Journal_Title :
Communications, IET
Publisher :
iet
ISSN :
1751-8628
Type :
jour
DOI :
10.1049/iet-com:20070046