• DocumentCode
    112097
  • Title

    Magnetically Actuated Soft Capsule With the Multimodal Drug Release Function

  • Author

    Sehyuk Yim ; Goyal, Keffy ; Sitti, Metin

  • Author_Institution
    Dept. of Mech. Eng., Carnegie Mellon Univ., Pittsburgh, PA, USA
  • Volume
    18
  • Issue
    4
  • fYear
    2013
  • fDate
    Aug. 2013
  • Firstpage
    1413
  • Lastpage
    1418
  • Abstract
    In this paper, we present a magnetically actuated multimodal drug release mechanism using a tetherless soft capsule endoscope for the treatment of gastric disease. Because the designed capsule has a drug chamber between both magnetic heads, if it is compressed by the external magnetic field, the capsule could release a drug in a specific position locally. The capsule is designed to release a drug in two modes according to the situation. In the first mode, a small amount of drug is continuously released by a series of pulse type magnetic field (0.01-0.03 T). The experimental results show that the drug release can be controlled by the frequency of the external magnetic pulse. In the second mode, about 800 mm3 of drug is released by the external magnetic field of 0.07 T, which induces a stronger magnetic attraction than the critical force for capsule´s collapsing. As a result, a polymeric coating is formed around the capsule. The coated area is dependent on the drug viscosity. This paper presents simulations and various experiments to evaluate the magnetically actuated multimodal drug release capability. The proposed soft capsules could be used as minimally invasive tetherless medical devices with therapeutic capability for the next generation capsule endoscopy.
  • Keywords
    diseases; drug delivery systems; drugs; endoscopes; magnetic actuators; magnetic fields; magnetic heads; medical robotics; polymer films; viscosity; capsule collapsing critical force; drug chamber; drug viscosity; external magnetic field; external magnetic pulse frequency; gastric disease treatment; magnetic attraction; magnetic flux density 0.01 T to 0.03 T; magnetic flux density 0.07 T; magnetic heads; magnetically actuated multimodal drug release capability evaluation; magnetically actuated soft capsule; minimally invasive tetherless medical devices; next generation capsule endoscopy; polymeric coating; pulse type magnetic field series; tetherless soft capsule endoscope; therapeutic capability; Coatings; Drugs; Endoscopes; Force; Magnetic resonance imaging; Magnetometers; Polymers; Capsule endoscope; drug delivery; magnetic micro-robot; medical robotics;
  • fLanguage
    English
  • Journal_Title
    Mechatronics, IEEE/ASME Transactions on
  • Publisher
    ieee
  • ISSN
    1083-4435
  • Type

    jour

  • DOI
    10.1109/TMECH.2012.2235077
  • Filename
    6401195