DocumentCode :
1129632
Title :
Single coil pair transcutaneous energy and data transceiver for low power bio-implant use
Author :
Vilches, Antonio ; Sanni, A. ; Toumazou, Christofer
Author_Institution :
Inst. of Biomed. Eng., Imperial Coll. London, London
Volume :
45
Issue :
14
fYear :
2009
Firstpage :
727
Lastpage :
728
Abstract :
A transcutaneous energy and data transceiver built from low-cost off-the-shelf components, employing a single pair of planar air-cored coils fabricated on standard PCB substrates, is presented. The transceiver powers the receiver continuously via the parallel-tuned coil pair. The latter is then able to sample audio at 8000 bytes/s and send the bitstream to a digital-to-analogue convertor in the transmitter by amplitude modulating the transmitted power carrier signal.
Keywords :
amplitude modulation; biomedical electronics; digital-analogue conversion; low-power electronics; prosthetics; transceivers; amplitude modulation; bit stream; byte rate 8000 Byte/s; data transceiver; digital-to-analogue convertor; low power bio-implant; parallel-tuned coil pair; planar air-cored coils; transcutaneous energy; transmitter;