DocumentCode
113145
Title
Setup and Validation of an MPI Signal Chain for a Drive Field Frequency of 150 kHz
Author
Sattel, Timo F. ; Woywode, Oliver ; Weizenecker, Jurgen ; Rahmer, Jurgen ; Gleich, Bernhard ; Borgert, Jorn
Author_Institution
Philips Med. Syst. DMC GmbH, Hamburg, Germany
Volume
51
Issue
2
fYear
2015
fDate
Feb. 2015
Firstpage
1
Lastpage
3
Abstract
In this paper, an imaging system is considered, which is part of an activity to test the feasibility of clinical (whole-body) magnetic particle imaging (MPI). Recent studies on nerve stimulation in humans indicate that drive field amplitudes have to be limited to lower values compared with preclinical MPI. Since nerve stimulation thresholds increase with frequency, the drive field frequency for the clinical demonstrator is set to 150 kHz. To date, available MPI systems usually apply drive fields at frequencies ranging from 1 to 25 kHz. We report on the technical feasibility of a signal chain setup for a drive field frequency of 150 kHz.
Keywords
biomedical equipment; biomedical imaging; magnetic particles; MPI signal chain; clinical demonstrator; clinical magnetic particle imaging; drive field amplitudes; drive field frequency; frequency 150 kHz; imaging system; nerve stimulation thresholds; Band-pass filters; Capacitors; Coils; Damping; Harmonic analysis; Imaging; Magnetic particles; Magnetic particle imaging (MPI); signal chain;
fLanguage
English