DocumentCode :
1132012
Title :
Novel Digital Voltage Ramp Generator for Use in Precision Current Sources in the Picoampere Range
Author :
Willenberg, Gerd-Dietmar ; Tauscher, H. Norbert
Author_Institution :
Phys.-Tech. Bundesanstalt, Braunschweig
Volume :
58
Issue :
4
fYear :
2009
fDate :
4/1/2009 12:00:00 AM
Firstpage :
756
Lastpage :
760
Abstract :
We report on the development of a novel highly linear voltage ramp generator to be used for the traceable calibration of picoamperemeters. The generator is based on two digital-to-analog converters, one of them being used to compensate for the differential nonlinearity of the other one. The generator is completely controllable by a computer, its voltage slope dV/dt is adjustable between 1 and 1000 mV/s. During a ramp running between -10 and +10 V, the slope shows a relative variation of only 1.3 middot10-5 (relative standard deviation). Due to a small output filter time constant of only 10 ms, we are, for the first time, able to conduct dynamic measurements of picoamperemeters with no sacrifice of linearity.
Keywords :
ammeters; calibration; digital-analogue conversion; electric current measurement; signal generators; differential nonlinearity; digital voltage ramp generator; digital-to-analog converters; picoamperemeter traceable calibration; precision current sources; Ammeters; calibration; current measurement; digital–analog conversion; signal generators;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2008.2010675
Filename :
4768695
Link To Document :
بازگشت