DocumentCode
1135448
Title
Cryogenic current comparator bridge for quantum Hall resistance ratio measurements
Author
Kleinschmidt, P. ; Williams, J.M. ; Fletcher, N.E. ; Janssen, T.J.B.M.
Author_Institution
Nat. Phys. Lab., Teddington, UK
Volume
149
Issue
6
fYear
2002
Firstpage
302
Lastpage
304
Abstract
A cryogenic current comparator (CCC) bridge for quantum Hall resistance ratio measurement has been constructed. Compared to previous CCC systems at the National Physical Laboratory, UK, the new system differs in two aspects. Firstly, it uses a type-II CCC, i.e. the ratio windings are on the outside and the detection winding is on the inside. This has the advantage that the CCC can be made very compact and fits into a standard storage Dewar. Secondly, the bridge electronics uses a digital servo and the SQUID feedback loop is disabled during current reversals. The overall relative measurement uncertainty of the system is of the order of 10-9.
Keywords
SQUIDs; bridge instruments; circuit feedback; cryogenics; current comparators; electric resistance measurement; measurement standards; measurement uncertainty; quantum Hall effect; servomechanisms; windings; He; SQUID feedback loop; cryogenic current comparator bridge; current reversals; digital servo bridge electronics; exterior ratio windings; interior detection winding; quantum Hall effect; quantum Hall resistance ratio measurements; relative system measurement uncertainty; storage Dewar fitting device; superconducting quantum interference devices; type-II CCC bridge;
fLanguage
English
Journal_Title
Science, Measurement and Technology, IEE Proceedings -
Publisher
iet
ISSN
1350-2344
Type
jour
DOI
10.1049/ip-smt:20020758
Filename
1176534
Link To Document