DocumentCode
1135455
Title
Design and construction of a 5 T 1028 mm bore magnetic resonance imaging (MRI) magnet
Author
Wang, S.T. ; Wahrer, R. ; Su, M. ; Juang, Y.F. ; Anet, F. ; Chen, C.J. ; Jiang, H.
Author_Institution
Wang NMR Inc., Livermore, CA, USA
Volume
30
Issue
4
fYear
1994
fDate
7/1/1994 12:00:00 AM
Firstpage
2348
Lastpage
2351
Abstract
New York Neurological Institute of Columbia University, requires a 5 T 1028 mm bore MRI magnet for brain function imaging and spectroscopy use. The magnet has a persistency of about 3×10-8/hr, a field uniformity of 0.4 ppm over 40 cm DSV and a liquid helium boil off of 0.55 litres per hour. In order to achieve good field uniformity, 12th order field uniformity coil windings was constructed. In addition, 18 sets of superconducting shim coils are built for field correction. The magnet weighs 26 tons. Design and construction will be described
Keywords
biomedical NMR; electromagnets; magnetic resonance spectrometers; superconducting magnets; 5 T; 5 T 1028 mm bore; MRI magnet; field correction; field uniformity; magnetic resonance imaging magnet; persistency; superconducting shim coils; Boring; Employment; Helium; Magnetic resonance imaging; Nitrogen; Nuclear magnetic resonance; Spectroscopy; Superconducting coils; Superconducting magnets; Temperature;