• DocumentCode
    1135455
  • Title

    Design and construction of a 5 T 1028 mm bore magnetic resonance imaging (MRI) magnet

  • Author

    Wang, S.T. ; Wahrer, R. ; Su, M. ; Juang, Y.F. ; Anet, F. ; Chen, C.J. ; Jiang, H.

  • Author_Institution
    Wang NMR Inc., Livermore, CA, USA
  • Volume
    30
  • Issue
    4
  • fYear
    1994
  • fDate
    7/1/1994 12:00:00 AM
  • Firstpage
    2348
  • Lastpage
    2351
  • Abstract
    New York Neurological Institute of Columbia University, requires a 5 T 1028 mm bore MRI magnet for brain function imaging and spectroscopy use. The magnet has a persistency of about 3×10-8/hr, a field uniformity of 0.4 ppm over 40 cm DSV and a liquid helium boil off of 0.55 litres per hour. In order to achieve good field uniformity, 12th order field uniformity coil windings was constructed. In addition, 18 sets of superconducting shim coils are built for field correction. The magnet weighs 26 tons. Design and construction will be described
  • Keywords
    biomedical NMR; electromagnets; magnetic resonance spectrometers; superconducting magnets; 5 T; 5 T 1028 mm bore; MRI magnet; field correction; field uniformity; magnetic resonance imaging magnet; persistency; superconducting shim coils; Boring; Employment; Helium; Magnetic resonance imaging; Nitrogen; Nuclear magnetic resonance; Spectroscopy; Superconducting coils; Superconducting magnets; Temperature;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.305747
  • Filename
    305747