Title
Design and construction of a 9.38 T, 183 mm-bore NMR/MRI spectrometer magnet
Author
Wang, S.T. ; Wahrer, R. ; Juang, Y.F. ; Chen, C.J. ; Su, M.
Author_Institution
Wang NMR Inc., Livermore, CA, USA
Volume
30
Issue
4
fYear
1994
fDate
7/1/1994 12:00:00 AM
Firstpage
2581
Lastpage
2584
Abstract
A 9.38 T NMR/MRI spectrometer magnet with a 183 mm warm bore has been designed, built and tested. The field uniformity requirements are 0.6 ppm over 25 mm diameter, 2.0 ppm for 40 mm diameter and 12 ppm for 80 mm diameter. In order to assure the achievement of the field uniformity, a set of 18 superconducting shim coils was built. In addition, another set of 18 resistive shim coils was built to correct the diamagnetism of the specimen. The results of a performance test for these systems are described
Keywords
biological NMR; biological techniques and instruments; magnetic resonance spectrometers; nuclear magnetic resonance imaging; spectrometer components and accessories; superconducting magnets; 183 mm; 25 mm; 40 mm; 80 mm; 9.38 T; NMR/MRI spectrometer magnet; diamagnetism correction; field uniformity; magnetic resonance imaging; neurological research; nuclear magnetic resonance; performance test; resistive shim coils; superconducting shim coils; warm bore; Boring; Conductors; Helium; Inductance; Magnetic resonance imaging; Nitrogen; Nuclear magnetic resonance; Spectroscopy; Superconducting coils; Superconducting magnets;