DocumentCode :
1138426
Title :
Neutron detection experiments with ionization counter and 6Li converter
Author :
Engels, R. ; Clemens, U. ; Kemmerling, G. ; Schelten, J.
Author_Institution :
Forschungszentrum Juelich GmbH, Germany
Volume :
51
Issue :
5
fYear :
2004
Firstpage :
2131
Lastpage :
2134
Abstract :
The detection of neutrons with thin 6Li containing films or layers within an ionization chamber is discussed. A detection efficiency of 50% for thermal neutrons is obtainable with three 100-μm thin 6Li foils. Preliminary measurements were performed with converters which contain a 1-μm thin 6LiF layer. The detector efficiency is only 3%. Pulse height distributions were measured and analyzed for the gases Ar, CF4, and C2F6 at pressures between 0.5 and 5 atm and at voltages between 50 and 500 V. Diffraction peaks from polycrystalline iron and semi-crystalline pyrolytic graphite were measured in TOF mode and are presented.
Keywords :
argon; convertors; graphite; ionisation chambers; iron; lithium; lithium compounds; neutron detection; organic compounds; pulse height analysers; scintillation counters; thin films; 0.5 to 5 atm; 1 micron; 100 micron; 50 to 500 V; 6Li converter; Ar; C; Fe; LiF; TOF mode; carbon tetrafluoride; detection efficiency; diffraction peaks; hexa fluoroethylene; ionization chamber; ionization counter; neutron detection experiments; polycrystalline iron; preliminary measurements; pulse height distributions; semicrystalline pyrolytic graphite; thermal neutrons; thin 6Li foils; thin 6LiF layer; Argon; Counting circuits; Detectors; Diffraction; Gases; Ionization chambers; Neutrons; Performance evaluation; Pulse measurements; Voltage;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2004.836040
Filename :
1344296
Link To Document :