• DocumentCode
    1141596
  • Title

    Repetitive waveform high frequency, high precision digitizer

  • Author

    Gyles, Colin

  • Author_Institution
    Analogic Corp., Danvers, MA, USA
  • Volume
    38
  • Issue
    4
  • fYear
    1989
  • fDate
    8/1/1989 12:00:00 AM
  • Firstpage
    917
  • Lastpage
    920
  • Abstract
    A digitizing technique for measuring high-speed phenomena with very high precision is described. A commercially available waveform digitizer using this method is shown to exhibit settling times of less than 5 ns to 0.1% and less than 10 ns to 0.01% after a voltage transition. Also, RMS measurements of sine waves up to 100 MHz show errors on the order of that obtained using a thermal transfer standard
  • Keywords
    analogue-digital conversion; wave analysers; 10 ns; 100 MHz; 5 ns; high precision digitizer; high-speed phenomena; repetitive waveform digitiser; sine waves; waveform analyser; Analog-digital conversion; Counting circuits; Delay effects; Frequency; Measurement standards; Nanoscale devices; Operational amplifiers; Time factors; Time measurement; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.31014
  • Filename
    31014