• DocumentCode
    1142966
  • Title

    Collimator Effects in High Resolution X-Ray Computed Tomography

  • Author

    Verly, Jacques G.

  • Volume
    1
  • Issue
    2
  • fYear
    1982
  • Firstpage
    122
  • Lastpage
    136
  • Abstract
    This paper complements other studies in the field of image restoration in X-ray computed tomography by including the source collimator among the sources of image degradation. The overall blur is describable in terms of a space variant 2-D impulse response or point spread function, which is initially calculated for an arbitrary scanning eccentricity, a uniform source intensity distribution, a uniform detector sensitivity, and arbitrary collimator aperture and position, and subsequently derived for more complicated sources and detectors. Numerous 3-D hidden-line displays are used throughout to illustrate the complicated 2-D functions emerging from the analysis.
  • Keywords
    Apertures; Collimators; Computed tomography; Degradation; Detectors; Image restoration; Optical imaging; Strips; X-ray detection; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1982.4307559
  • Filename
    4307559