    Development of a GSO detector assembly for a continuous blood sampling system

  • Author

    Kudomi, Nobuyuki ; Choi, Eunjoo ; Yamamoto, Seiichi ; Watabe, Hiroshi ; Kim, Kyeong Min ; Shidahara, Miho ; Ogawa, Mikako ; Teramoto, Noboru ; Sakamoto, Eishi ; Iida, Hidehiro

  • Author_Institution
    Nat. Cardiovascular Center Res. Inst., Osaka, Japan
  • Volume
    50
  • Issue
    1
  • fYear
    2003
  • fDate
    2/1/2003 12:00:00 AM
  • Firstpage
    70
  • Lastpage
    73
  • Abstract
    A new input function monitoring system has been developed and evaluated using a GSO detector assembly for both PET and SPECT quantitative studies. Energy resolutions were 11% for 511 keV photons, 20% for 140 keV(99mTc) photons and 28% for 70 keV(201Tl) photons, enabling the use of this system in SPECT studies. The paired assembly of crystals provided an absolute sensitivity of approximately 7% for PET tracers and 70% for 99mTc and 201Tl (SPECT tracers). Multiple arrangement of paired detectors would make it possible to correct for the transit time of radioactivity through the catheter tube. This study demonstrates that the present system can be of use in both clinical and small animal studies using SPECT and PET tracers.
  • Keywords
    biomedical equipment; positron emission tomography; single photon emission computed tomography; solid scintillation detectors; 140 keV; 511 keV; 70 keV; 201Tl; 99mTc; GSO detector; Gd2SiO5; PET; SPECT; Tc; Tl; blood sampling system; catheter; energy resolution; small animal; Assembly systems; Blood; Catheters; Detectors; Energy resolution; Monitoring; Photonic crystals; Positron emission tomography; Sampling methods; Single photon emission computed tomography;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2002.807869
  • Filename
    1178692