• DocumentCode
    1143846
  • Title

    Spatial Resolution in NMR Imaging

  • Author

    Crooks, Lawrence E. ; Kaufman, Leon ; Hoenninger, John, III ; Arakawa, Mitsuaki ; Watts, Jeffrey ; Cannon, Clifford R.

  • Volume
    3
  • Issue
    2
  • fYear
    1984
  • fDate
    6/1/1984 12:00:00 AM
  • Firstpage
    51
  • Lastpage
    53
  • Abstract
    NMR imaging is used as an example of how spatial resolution can be improved in a signal-to-noise (S/N) limited situation. The NMR imaging process consists of two components-generating the NMR signal and localizing it in space. This paper will show that spatial resolution not only aids in identifying small structures, but improves the detectability of larger features by preserving their object contrast.
  • Keywords
    Computer vision; Image reconstruction; Magnetic fields; Magnetic resonance imaging; Nuclear magnetic resonance; Object detection; Radio frequency; Signal generators; Signal processing; Spatial resolution;