DocumentCode :
1144201
Title :
Finite-Element Analysis and In Vitro Experiments of Placement Configurations Using Triple Antennas in Microwave Hepatic Ablation
Author :
Phasukkit, Pattarapong ; Tungjitkusolmun, Supan ; Sangworasil, Manas
Author_Institution :
Dept. of Electron. & Comput. Service Center, King Mongkut´´s Inst. of Technol. Ladkrabang, Bangkok, Thailand
Volume :
56
Issue :
11
fYear :
2009
Firstpage :
2564
Lastpage :
2572
Abstract :
This study presents analyses of triple-antenna configurations and designs for microwave (MW) hepatic ablation using 3-D finite-element (FE) analyses verified by in vitro experiments. Treatment of hepatic cancer often requires removal or destruction of large volume lesions. Using multiple antennas offers a potential solution for creating ablation zones with larger dimensions, as well as varied geometrical shapes. We performed both 3-D FE analyses and in vitro experiments using three identical open-tip MW antennas simultaneously, placing them in three types of configurations-ldquolinear array,rdquo ldquotriangular,rdquo and ldquoT-shapedrdquo arrangements. We compared coagulation volumes created, as well as temperature distribution characteristics, from the three-antenna arrangements after power delivery of 50 W for 60 s. We also performed additional tests using nonidentical antennas (open tip, slot, and slot with insulating jacket) for the three configurations. The results illustrate that arranging antennas in the ldquoT-shapedrdquo pattern destroyed more unwanted tissues than those found when using ldquolinear arrayrdquo and ldquotriangularrdquo arrangements, with maximum coagulation width and depth of 46 and 81 mm, respectively, and coagulation volume of 30.7 cm3 . In addition, using nonidentical triple antennas caused variations in coagulation zone characteristics, and thus, the technique could be applied to treatment situations where nonsymmetric coagulation zones are required.
Keywords :
antenna arrays; cancer; finite element analysis; liver; microwave heating; 3-D finite-element analyses; coagulation volumes; finite-element analysis; hepatic cancer; microwave hepatic ablation; temperature distribution; triple antennas; Antenna arrays; Cancer; Coagulation; Finite element methods; In vitro; Iron; Lesions; Load flow; Load flow analysis; Shape; Bioheat equation; finite-element (FE) analysis; hepatic ablation; microwave (MW) ablation; triple antenna; Ablation Techniques; Animals; Computer Simulation; Equipment Design; Finite Element Analysis; Liver; Liver Neoplasms; Microwaves; Swine;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2009.2027128
Filename :
5170082
Link To Document :
بازگشت