DocumentCode :
1144346
Title :
Projection Space Iteration Reconstruction-Reprojection
Author :
Kim, J.H. ; Kwak, K.Y. ; Park, S.-B. ; Cho, Z.H.
Volume :
4
Issue :
3
fYear :
1985
Firstpage :
139
Lastpage :
143
Abstract :
Recently, an iterative reconstruction-reprojection (IRR) algorithm has been suggested for application to limited view computed tomography (CT). In the IRR, the interpolation operation is performed in the object space during backprojection-reprojection. The errors associated with the interpolation degrade the reconstructed image and may cause divergence unless a large number of rays is used. In this paper, we propose a projection space iterative reconstruction-reprojection (PSIRR) algorithm based on backprojection-reprojection in the projection space. In the process of the backprojection-reprojection, iteration is performed with a single equation instead of multiple equations and interpolation is eliminated. Computer simulation results are presented, and image quality of the PSIRR algorithm shows a substantial improvement compared to the original IRR algorithm. In addition, the new algorithm was applied to ultrasonic attenuation CT using a sponge phantom.
Keywords :
Attenuation; Computed tomography; Computer simulation; Degradation; Equations; Image quality; Image reconstruction; Imaging phantoms; Interpolation; Iterative algorithms;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.1985.4307712
Filename :
4307712
Link To Document :
بازگشت